The Gamma Counter LG-4 is designed for radioimmunoassay (RIA) and immunoradiometric assay (IRMA) in medium size laboratories employing isotope 125I. The counter is an easy to operate instrument with no moving mechanical components enlarging his operation reliability.
The gamma counter carries out measurement up to 4 samples in 4 measuring channels. 125I
radiation is detected with NAI(Tl) well scintillators ensuring high detection efficiency of gamma radiation.
Automatic gain control of photomultiplier tube for each measuring channel, based on light pulse from light
emitting diode as reference, ensures high stability of the counter operation. Hardcopy of the calibration
curve and of the measuring results is made on a band paper by a needle printer. Lead shielding around the radiation
detectors and registration of the detector pulses in a window corresponding to 125I energy range results in
low background of natural radiation ensuring thus measurement of samples with low substance concentration. |
| Type of isotope measured | 125I |
| Number of detectors | 4 |
| Radiation detectors | NaI(Tl) ø30x40mm total dimension, ø17x38mm well |
| Calibration curve RIA | lin(B/B0) versus log(CONC) |
| Calibration curve IRMA | log(B/Bmax) versus log(CONC) |
| Measuring time | 1 ÷ 10 min |
| Gain control | automatic, LED light pulse as reference |
| External printer | EPSON LX-300+ |
| Power supply | 220 V, +10%, -15% |
| Serial port | RS232C |
| Dimensions | 310x220x300mm |
| Weight | approx. 25 kg |